And now, for something completely different... By the way, it is a continuation of the previous article. I'm sorry I ended up halfway last time. Please be assured that this article will conclude.
Previous article: Electron microscope simulation with Python: Multislice method (1)
The code is GitHub [^ 1]
Note again: If you are simulating as a hobby (hereinafter referred to as hobby ration), it is difficult to determine whether the result is really correct, and the result shown here may not be physically correct. I would appreciate it if you could point out any mistakes.
At the end of the previous article, I said that multi-slice calculations alone are not enough to simulate crystal structures. Specifically, what is lacking is that the multi-slice calculation alone does not take into account the difference from the ideal system due to the nature of the electron microscope, which is the observation instrument. It is not limited to electron microscopes, but when simulating measurements, the purpose is mostly to analyze the measurement results. Therefore, the simulation needs to be as close to the measurement system as possible. There are many differences (aberrations) from the ideal system that can be considered with an electron microscope. Among them, spherical aberration and chromatic aberration are particularly effective for the crystal structure image by HREM.
Spherical aberration is an aberration caused by the objective lens of an electron microscope. When an electron beam enters the objective lens, the electron beam incident at an angle close to perpendicular to the objective lens (parallel to the optical axis) properly gathers on the image plane, but the electron beam incident at an angle from the optical axis The focus will be on the distance from the image plane. The blurring of the image due to this is spherical aberration. The magnitude of spherical aberration changes depending on the performance of the objective lens.
Spherical aberration is expressed by the following formula.
Considering that the phase contrast is reflected properly at $ CTF = -0.5 $, it can be said that if $ \ alpha $ is around 0.0025 to 0.0125, it is within the acceptable range. Since the black and white of the image is inverted when the CTF positive or negative changes, the phase contrast will be messed up where the positive or negative of the CTF after $ \ alpha = 0.0125 $ fluctuates greatly.
Now that you know the CTF trends, let's plot the CTF in the range calculated in the previous article.
Considering only spherical aberration, it was found that the phase contrast seems to be correct in almost the entire range.
Chromatic aberration is another factor that has a large effect on the crystal structure image. Color refers to wavelength. Chromatic aberration is due to the width of the electron beam energy. The difference in wave energy is the difference in wavelength, so it is the difference in color. The electron beam emitted from the electron gun of the electron microscope has a spread (fluctuation) of about $ \ Delta E / E = 10 ^ {-5} $ or less. Furthermore, the current of the lens that collects the electron beam also has the same fluctuation $ \ Delta J / J
Let's look at the previous calculation range as before.
If you make a crystal structure image based on these, you should be able to make a more correct image.
Spherical aberration and chromatic aberration are calculated in advance and convolved into $ \ Psi_ {out} $ output from the multi-slice calculation.
Cs = 0.5e-3
deltaf = 1.2*(Cs*lamb)**(1/2)
hkl = [h, k, 0]
thkl = np.transpose(hkl)
dk = 1/((np.matmul(np.matmul(invG, thkl), hkl))**(1/2))
u = self.lamb/(2*dk)
chi = 2*np.pi/lamb
chi = chi*(1/4*self.Cs*u**4 - 1/2*deltaf*u**2)
chromatic aberration
deltaE = 1.0e-6
deltaJ = 0.5e-6
Cc = 1.4e-3
sig = Cc*((deltaE)**2 + (2*deltaJ)**2)**(1/2)
hkl = [h, k, 0]
thkl = np.transpose(hkl)
dk = 1/((np.matmul(np.matmul(invG, thkl), hkl))**(1/2))
u = lamb/(2*dk)
w = np.exp(-(1/2)*(np.pi**2)*(u**4)*(sig**2)/(lamb**2))
α-Fe, [001] Incident, accelerating voltage 200 keV, $ C_S $ = 0.5 mm, $ C_C $ = 1.4 mm, $ \ Delta E / E $ = 1.0 μm, $ \ Delta J / J $ = 0.5 μm Then, the crystal structure image looks like this.
When $ C_S $ is set to 0.5 μm, it becomes as follows. The image is clearer. A microscope with a small aberration coefficient is a good microscope.
It's been a long time, but I made a simulation of the crystal structure image with HREM. Actually, there are some parameters such as convergence angle that have not been taken into consideration yet, so I hope to improve them in the future.